Medtronic MiniMed recalls MiniMed Go app version 1.2.1 because the long-acting dose reminder can stop working
After updating to version 1.2.1, users who had turned on the long-acting dose reminder in version 1.2.0 or earlier may get no reminder, which could lead to missed or delayed insulin doses and high glucose, or double dosing and low glucose.
What to do
The notice does not give instructions. Read the official notice for details.
Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.
Recall details
- Product
- Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app for iOS or MiniMed Go app for Android is compatible with the following components: " Simplera sensor " Instinct sensor " InPen smart insulin pen " Apple Watch [Optional] Continuous glucose monitoring…
- Company
- Medtronic MiniMed, Inc.
- Hazard
- A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new…
- Units
- 623
- Where sold
- Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom.
- FDA classification
- Class II · Ongoing
- Classified by FDA
- October 22, 2026
- Recall started
- September 2, 2026
- FDA enforcement report
- October 28, 2026
- Identifiers
- Codes: Lot Code: GTIN for MMT-8600, iOS: 00199150038978 GTIN for MMT-8601, Android: 00199150038985 UDI for MMT-8600, iOS: 00199150038978 UDI for MMT-8601, Android: 00199150038985 Software version 1.2.1
- FDA recall number: Z-0265-2027
The headline, summary and steps above are written in plain English from the official notice below. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.
Official notice
The notice's own wording from FDA.
- Product
- Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app for iOS or MiniMed Go app for Android is compatible with the following components: " Simplera sensor " Instinct sensor " InPen smart insulin pen " Apple Watch [Optional] Continuous glucose monitoring (CGM) is a technology that allows you to continuously view glucose values. The app uses an inserted Simplera sensor CGM or Instinct sensor CGM with Bluetooth¿ wireless technology that converts small amounts of glucose from the interstitial fluid under the skin into an electronic signal. The sensors use that signal to provide sensor glucose (SG) values to the MiniMed Go app on a compatible mobile device. The app also provides alerts based on your SG levels and contains a user interface for entering data such as insulin, meals, exercise, blood glucose (BG) values, and uploading information to the CareLink" Personal website. The app is available to download from the Apple"* App Store"* or the Google Play"* Store. Component: No
- Reason for recall
- A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.
- Classification
- Class II (Ongoing)
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom.
- Quantity
- 623
Related recalls
Medtronic MiniMed recalls CareLink Clinic software version 4.4A over incorrect insulin setting calculations
The Settings Calculator in CareLink Clinic software version 4.4A does not use custom values entered by healthcare professionals, so it gives incorrect rapid-acting insulin settings and a patient could get higher or lower insulin doses…
Medtronic MiniMed, Inc. · 1 unit
Medtronic MiniMed recalls MiniMed Flex insulin pumps over missing or failed test results
Some MiniMed Flex insulin pumps (MMT-8062 and MMT-8072) were released with missing or failed results in one or more of 9 functional tests, so they may not work at setup, charge slowly, or deliver insulin inaccurately.
Medtronic MiniMed, Inc. · 25 units
Medtronic MiniMed recalls MiniMed 780G, 670G and 640G insulin pumps over loaner pumps cleaned with wrong products
Loaner insulin pumps may have been cleaned with products that do not mix safely with them, which could damage the case or keypad, let moisture in, and cause too much or too little insulin, leading to hypoglycemia, hyperglycemia, or…
Medtronic MiniMed, Inc. · 280 units
Medtronic MiniMed recalls MiniMed Go App over alerts that may not sound at night
In MiniMed Go app versions 1.0.0 and 1.1.0, when Set day and night is turned OFF, missed dose or Correct high glucose alerts will not be triggered at night, which could mean missed rapid-acting insulin doses without the user knowing and a…
Medtronic MiniMed, Inc. · 3,465 units
Medtronic MiniMed recalls MiniMed 780G insulin pumps over software problem that can stop insulin delivery
When paired with the Instinct sensor, the pump can unexpectedly stop delivering insulin and show Pump Error 53 then Pump Error 23, which may delay treatment and raise the risk of hyperglycemia or diabetic ketoacidosis (DKA).
Medtronic MiniMed, Inc. · 778 units
Medtronic MiniMed recalls insulin pumps over too much or too little insulin when pump height changes
Insulin over-delivery when the pump is elevated can cause severe hypoglycemia, seizure, coma, or death, and under-delivery when lowered can cause severe hyperglycemia, diabetic ketoacidosis, or death.
Medtronic MiniMed, Inc. · 1,086,165 units