Medtronic MiniMed recalls MiniMed 780G, 670G and 640G insulin pumps over loaner pumps cleaned with wrong products
Loaner insulin pumps may have been cleaned with products that do not mix safely with them, which could damage the case or keypad, let moisture in, and cause too much or too little insulin, leading to hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
What to do
The notice does not give instructions. Read the official notice for details.
Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.
Recall details
- Product
- MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW; MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;; MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
- Company
- Medtronic MiniMed, Inc.
- Hazard
- Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
- Units
- 257; 7; 16
- Where sold
- International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
- FDA classification
- Class II · Ongoing
- Classified by FDA
- August 18, 2026
- Recall started
- July 13, 2026
- FDA enforcement report
- August 26, 2026
- Identifiers
- Codes: Model/UDI-DI Serial Numbers: MMT-1885CLW/00763000442187, 00763000974374/NG3150737H, NG3155194H, NG3151561H, NG3154693H, NG3151716H, NG3153547H, NG3153637H, NG3155407H, NG3152666H, NG3155788H, NG3081975H, NG2901117H, NG3080842H, NG3081463H, NG3153416H, NG3081894H, NG3153781H, NG3155944H, NG3081982H, NG3081525H, NG3155909H, NG3090722H, NG3154897H, NG3088436H, NG3088708H, NG3090073H, NG3155908H, NG3150720H, NG3092957H, NG3128715H, NG3128784H, NG3152774H, NG3153929H, NG3514566H, NG3150780H, NG3151290H, NG3151508H, NG3507161H, NG3508693H, NG3081976H, NG3151460H, NG3152051H, NG3152085H, NG3152344H, NG3153213H, NG3152772H, NG3155586H, NG3150786H, NG3153214H, NG3153385H, NG3151840H, NG3082019H…
- FDA recall number: Z-2982-2026
- Codes: Model/UDI-DI Serial Numbers: MMT-1781/00763000404536/NG2665223H, NG2748200H, NG2748664H, NG2755237H, NG2716176H, NG3547582H. MMT-1782/00763000404543/NG2768190H
- FDA recall number: Z-2981-2026
- Codes: Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H
- FDA recall number: Z-2980-2026
The headline, summary and steps above are written in plain English from the official notice below. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.
Official notice
The notice's own wording from FDA.
- Product
- MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW | MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782; | MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
- Reason for recall
- Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
- Classification
- Class II (Ongoing)
- Distribution
- International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
- Quantity
- 257; 7; 16
Related recalls
Medtronic MiniMed recalls MiniMed Go app version 1.2.1 because the long-acting dose reminder can stop working
After updating to version 1.2.1, users who had turned on the long-acting dose reminder in version 1.2.0 or earlier may get no reminder, which could lead to missed or delayed insulin doses and high glucose, or double dosing and low glucose.
Medtronic MiniMed, Inc. · 623 units
Medtronic MiniMed recalls CareLink Clinic software version 4.4A over incorrect insulin setting calculations
The Settings Calculator in CareLink Clinic software version 4.4A does not use custom values entered by healthcare professionals, so it gives incorrect rapid-acting insulin settings and a patient could get higher or lower insulin doses…
Medtronic MiniMed, Inc. · 1 unit
Medtronic MiniMed recalls MiniMed Flex insulin pumps over missing or failed test results
Some MiniMed Flex insulin pumps (MMT-8062 and MMT-8072) were released with missing or failed results in one or more of 9 functional tests, so they may not work at setup, charge slowly, or deliver insulin inaccurately.
Medtronic MiniMed, Inc. · 25 units
Medtronic MiniMed recalls MiniMed Go App over alerts that may not sound at night
In MiniMed Go app versions 1.0.0 and 1.1.0, when Set day and night is turned OFF, missed dose or Correct high glucose alerts will not be triggered at night, which could mean missed rapid-acting insulin doses without the user knowing and a…
Medtronic MiniMed, Inc. · 3,465 units
Medtronic MiniMed recalls MiniMed 780G insulin pumps over software problem that can stop insulin delivery
When paired with the Instinct sensor, the pump can unexpectedly stop delivering insulin and show Pump Error 53 then Pump Error 23, which may delay treatment and raise the risk of hyperglycemia or diabetic ketoacidosis (DKA).
Medtronic MiniMed, Inc. · 778 units
Medtronic MiniMed recalls insulin pumps over too much or too little insulin when pump height changes
Insulin over-delivery when the pump is elevated can cause severe hypoglycemia, seizure, coma, or death, and under-delivery when lowered can cause severe hyperglycemia, diabetic ketoacidosis, or death.
Medtronic MiniMed, Inc. · 1,086,165 units