Medtronic MiniMed recalls MiniMed Flex insulin pumps over missing or failed test results
Some MiniMed Flex insulin pumps (MMT-8062 and MMT-8072) were released with missing or failed results in one or more of 9 functional tests, so they may not work at setup, charge slowly, or deliver insulin inaccurately.
What to do
The notice does not give instructions. Read the official notice for details.
Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.
Recall details
- Product
- Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with…
- Company
- Medtronic MiniMed, Inc.
- Hazard
- Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump…
- Units
- 25
- Where sold
- U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.
- FDA classification
- Class II · Ongoing
- Classified by FDA
- October 5, 2026
- Recall started
- August 20, 2026
- FDA enforcement report
- October 14, 2026
- Identifiers
- Codes: MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB00015793, FB00015794, FB00015796, FB00015799, FB00015801, FB00015977, FB00017271. UDI-DI/GTIN: 00763000848156 Note: Initially 25 pumps were identified as put into service, however 2 were shipped to Marketing Team Members and recovered. 3 Pumps were returned through complaint process. Final total of pumps available in the field is 20. MMT-8062 Lot Code: Serial numbers: FB00011770, FB00015162, FB00016429, FB00014760, FB00015797 UDI-DI/GTIN: 00763000848149 (Serial Numbers: FB00017271, FB00011770…
- FDA recall number: Z-0088-2027
The headline, summary and steps above are written in plain English from the official notice below. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.
Official notice
The notice's own wording from FDA.
- Product
- Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
- Reason for recall
- Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
- Classification
- Class II (Ongoing)
- Distribution
- U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.
- Quantity
- 25
Related recalls
Medtronic MiniMed recalls MiniMed Go app version 1.2.1 because the long-acting dose reminder can stop working
After updating to version 1.2.1, users who had turned on the long-acting dose reminder in version 1.2.0 or earlier may get no reminder, which could lead to missed or delayed insulin doses and high glucose, or double dosing and low glucose.
Medtronic MiniMed, Inc. · 623 units
Medtronic MiniMed recalls CareLink Clinic software version 4.4A over incorrect insulin setting calculations
The Settings Calculator in CareLink Clinic software version 4.4A does not use custom values entered by healthcare professionals, so it gives incorrect rapid-acting insulin settings and a patient could get higher or lower insulin doses…
Medtronic MiniMed, Inc. · 1 unit
Medtronic MiniMed recalls MiniMed 780G, 670G and 640G insulin pumps over loaner pumps cleaned with wrong products
Loaner insulin pumps may have been cleaned with products that do not mix safely with them, which could damage the case or keypad, let moisture in, and cause too much or too little insulin, leading to hypoglycemia, hyperglycemia, or…
Medtronic MiniMed, Inc. · 280 units
Medtronic MiniMed recalls MiniMed Go App over alerts that may not sound at night
In MiniMed Go app versions 1.0.0 and 1.1.0, when Set day and night is turned OFF, missed dose or Correct high glucose alerts will not be triggered at night, which could mean missed rapid-acting insulin doses without the user knowing and a…
Medtronic MiniMed, Inc. · 3,465 units
Medtronic MiniMed recalls MiniMed 780G insulin pumps over software problem that can stop insulin delivery
When paired with the Instinct sensor, the pump can unexpectedly stop delivering insulin and show Pump Error 53 then Pump Error 23, which may delay treatment and raise the risk of hyperglycemia or diabetic ketoacidosis (DKA).
Medtronic MiniMed, Inc. · 778 units
Medtronic MiniMed recalls insulin pumps over too much or too little insulin when pump height changes
Insulin over-delivery when the pump is elevated can cause severe hypoglycemia, seizure, coma, or death, and under-delivery when lowered can cause severe hyperglycemia, diabetic ketoacidosis, or death.
Medtronic MiniMed, Inc. · 1,086,165 units