FDA·Medical devices·October 14, 2026

Medtronic MiniMed recalls MiniMed Flex insulin pumps over missing or failed test results

Some MiniMed Flex insulin pumps (MMT-8062 and MMT-8072) were released with missing or failed results in one or more of 9 functional tests, so they may not work at setup, charge slowly, or deliver insulin inaccurately.

What to do

The notice does not give instructions. Read the official notice for details.

Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.

Recall details

Product
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with…
Hazard
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump…
Units
25
Where sold
U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.
FDA classification
Class II · Ongoing
Classified by FDA
October 5, 2026
Recall started
August 20, 2026
FDA enforcement report
October 14, 2026
Identifiers
  • Codes: MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB00015793, FB00015794, FB00015796, FB00015799, FB00015801, FB00015977, FB00017271. UDI-DI/GTIN: 00763000848156 Note: Initially 25 pumps were identified as put into service, however 2 were shipped to Marketing Team Members and recovered. 3 Pumps were returned through complaint process. Final total of pumps available in the field is 20. MMT-8062 Lot Code: Serial numbers: FB00011770, FB00015162, FB00016429, FB00014760, FB00015797 UDI-DI/GTIN: 00763000848149 (Serial Numbers: FB00017271, FB00011770…
  • FDA recall number: Z-0088-2027

The headline, summary and steps above are written in plain English from the official notice below. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.

Official notice

The notice's own wording from FDA.

Product
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
Reason for recall
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
Classification
Class II (Ongoing)
Distribution
U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.
Quantity
25