Baxter recalls Likorall overhead lifts over strap that may lower faster than expected
A broken motor shaft combined with reduced performance of the Single Fault Safety (SFS) system may make the lift strap lower more quickly than expected during use, and patients are at some risk of injury if a significant fall or drop occurs.
What to do
The notice does not give instructions. Read the official notice for details.
Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.
Recall details
- Product
- Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Product Code: 3122007 Product Description…
- Hazard
- Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
- Where sold
- Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota…
- FDA classification
- Class II · Ongoing
- Classified by FDA
- October 13, 2026
- Recall started
- September 14, 2026
- FDA enforcement report
- October 21, 2026
- Identifiers
- Codes: Product Code: 3121001 Product Description: LIKORALL 200 Lot/Serial Number: Intervals: 8400054-8450499 and 9000000-9000999 UDI-DI Number: 00887761981911 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Lot/Serial Number: Greater than or equal to 8104548 UDI-DI Number: 00887761979710 Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Lot/Serial Number: Interval: 429087-431032 UDI-DI Number: 00887761979727 Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Lot/Serial Number: Greater than or equal to: 8101715 UDI-DI Number: 00887761979734 Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Lot/Serial Number: Interval…
- FDA recall number: Z-0151-2027
The headline, summary and steps above are written in plain English from the official notice below. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.
Official notice
The notice's own wording from FDA.
- Product
- Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122008NA Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122009 Product Description: LIKORALL 242 S, NATURAL Product Code: 3122009NA Product Description: LIKORALL 242 S, NATURAL Product Code: 3122010 Product Description: LIKORALL 242 S, WHITE Product Code: 3122011 Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122011NA Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122012 Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122012NA Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122501 Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122501NA Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122502 Product Description: LIKORALL 250 ES, WHITE Product Code: 3122502NA Product Description: LIKORALL 250 ES, WHITE Product Code: 3123001 Product Description: LIKORALL 243 ES Product Code: 3123001NA Product Description: LIKORALL 243 ES Product Code: 3123002 Product Description: LIKORALL 243 ES, WHITE Product Code: 3123002NA Product Description: LIKORALL 243 ES, WHITE Product Code: 3124050 Product Description: LIKORALL 250 S, IRC
- Reason for recall
- Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
- Classification
- Class II (Ongoing)
- Distribution
- Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota…
Related recalls
Baxter recalls PRO+ MRS hospital bed mattresses over labeling error that could lead to use on wrong bed frame
The product description on the mattresses' device identification label is wrong, which could lead to the wrong mattress being picked and used with a bed frame it does not fit.
Baxter Healthcare Corporation · 999 units
Baxter recalls VOLARA home care airway clearance system over possible flow rate error from test equipment contamination
Contamination in the factory's release test equipment could have made the nebulizer flow rate reading inaccurate, but testing of the returned devices confirmed they meet specifications.
Baxter Healthcare Corporation · 2 units
Baxter recalls 70% Dextrose Injection 2000 mL bags over stainless steel particles in the solution
Stainless steel particles were found in the solution in these intravenous bags, and the notice names this as the reason for the recall of 606 bags sold across the U.S.
Baxter Healthcare Corporation · 606 bags
Baxter recalls 0.9% Sodium Chloride Injection 500 mL bags over fiberglass particles
Some 500 mL bags of 0.9% Sodium Chloride Injection USP in VIAFLEX plastic containers were found to contain particles identified as fiberglass.
Baxter Healthcare Corporation · 145,464 500mL bags
Baxter recalls Novum IQ syringe pumps over software issue during multi-syringe infusions
A software problem can occur during a Volume/Time multi-syringe infusion when a syringe runs empty and is replaced while some volume still remains to be infused.
Baxter Healthcare Corporation · 14,605 units
Baxter recalls clindamycin, vasopressin, daptomycin and Cardene IV injections over manufacturing rule violations
The notice says these four Baxter premixed injection products were made with deviations from current good manufacturing practice, and it does not describe a specific defect or health risk.