FDA Class IFDA·Healthcare and labs·May 13, 2026

Ventec recalls VOCSN V+Pro ventilators over possible oxygen leak that could cause low oxygen delivery and fire risk

Production testing may not have covered all high-pressure conditions, so an oxygen leak could occur, giving patients less oxygen than intended and making fire more likely in the oxygen-enriched air.

What to do

The notice does not give instructions. Read the official notice for details.

Recall details

Product
VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
Hazard
Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
Units
42
Where sold
US: OH, GA, CA, MO
FDA classification
Class I · Ongoing
Classified by FDA
May 1, 2026
Recall started
March 25, 2026
FDA enforcement report
May 13, 2026
Identifiers
  • Codes: UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532
  • FDA recall number: Z-1942-2026

The headline, summary and steps above are written in plain English from the official notice below. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.

Official notice

The notice's own wording from FDA.

Product
VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000
Reason for recall
Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
Classification
Class I (Ongoing)
Distribution
US: OH, GA, CA, MO
Quantity
42